Here you can find the changelog of REALITi since it was posted on our website on 2025-05-10 16:04:24.
The latest version is 11.21.0 and it was updated on 2026-06-10 05:35:34. See below the changes in each version.
REALITi version 11.21.0
Updated At: 2026-06-10
Changes: Version History Atlas
Improved Pleth (SpO₂ waveform) behaviour during CPR so pulses remain visible during compressions and can reflect both intrinsic cardiac activity and CPR-generated pulses.
Introduced a dedicated Pulse Rate (PR) value separate from Heart Rate (HR), improving handling of pulseless states, CPR, low perfusion, and probe disconnection while keeping PR aligned with HR by default.
Camera
Exported videos now include ECG, SpO₂, and ETCO₂ waveforms as shown in live preview.
Corpuls
Improved AED analysis: removed incorrect “Patient Movement Detected” message and enhanced shock decision timing using ECG analysis blocks.
Added CPR depth support with manikin-specific ranges.
Improved CO₂ and apnoea alarm behaviour, including correct triggering and translations.
Added SpO₂ probe-off alarm with localized messaging.
Fixed metronome audio and defibrillator behaviour, including QRS marker display and power-on logic.
Ensured alarm notifications clear correctly after Autolimits adjustments.
CPR Band / CPR Puck
Improved Bluetooth connectivity and compatibility with newer devices.
CPR Console
Fixed compression depth unit conversion and display accuracy.
Facilitator/Control
Improved ECG waveform behaviour: heart rate now correctly updates or persists depending on waveform type.
Corrected passive rate values on select ECG waveforms and fixed a waveform naming typo.
Added configurable sound settings (volume, alarm, heartbeat) within events, including scenario saving, defaults, and clear display states.
Improved auscultation menu behaviour to prevent overlap and ensure proper closing during navigation.
Fixed layout issues with tab labels in languages with longer text.
Added collapsible folders in the Scenario Library.
Improved connection handling between Control and Monitor, including reconnection after Wi-Fi loss.
Resolved connection stability issues to prevent repeated reconnection attempts while awaiting monitor selection, eliminating device flickering.
Corrected unintended event behaviour so sound-related settings (e.g., Alarm Off) only appear when explicitly configured.
LIFEPAK 35
Corrected ETCO₂ sweep speed and waveform display to better match real device behaviour.
Menu / main screen
Fixed issue where Settings page could appear blank on some iOS devices.
Revel Vent
Fixed crash when connecting Control to the monitor.
SimConnect
Updated firmware (v1.4.0) with improved logging and behaviour consistency.
Improved Wi-Fi settings retrieval during SimConnect connection, including clearer loading prompts and retrying settings requests if the router is still starting up.
SimOccluder
Updated in-app branding.
Added support for SimOccluder firmware v1.2.0.
Zoll R
Added configuration options for AED startup mode and automatic AED analysis.
Zoll X
Fixed 12-lead ECG display reverting incorrectly in Torsades rhythm.
Added CPR metronome audio feedback.
Updated ETCO₂ monitoring to include automatic calibration on activation.
Multiple Screens
Fixed missing exit prompt on monitor screens.
Prevented automatic BP from starting while monitor is off.
Maintenance and General
Updated compatibility with latest Apple system updates.
Standardized screen orientation (landscape by default; portrait where required).
Fixed unit consistency in exported videos (ETCO₂ and temperature).
Fixed TestFlight launch crash. 11.21.0 Apr 30 SimOccluder
Released SimOccluder firmware version 1.1.0 to improve performance, reliability, and stability, including refined internal processes for a smoother and more consistent experience.
Updated the SimOccluder logo to reflect the latest branding.
SimConnect
Updated SimConnect firmware to version 1.4.0, adding Event Config logging to improve device logging and performance.
Atlas
Refreshed Adult, Junior, and Baby icons in the Atlas Connection Panel for improved clarity and consistency.
Released Atlas Baby firmware version 1.5.1 to enhance performance and reliability.
Added clearer pulse strength descriptions in manikin settings: 0 – Pulseless, 1 – Weak, 2 – Normal, 3 – Bounding.
Corpuls
Fixed an issue where the Analyse and Shock keys did not respond as expected; both now correctly open defibrillator mode.
Resolved an issue where C3 soft keys became inactive after switching from pacing to Monitor mode; controls now remain active.
Zoll X Advanced
Unified Advanced Settings for ZOLL X and ZOLL X Advanced into a single, consistent configuration view. Automatic NIBP and Medication List settings are now included within the shared experience.
Added a “Default Lead” setting to select Pads or a specific ECG Lead as the startup default across all monitor modes. Pads remain the default unless changed. If Pads are turned off, the monitor defaults to Lead II for ECG display.
Scenarios
Fixed an issue where shared scenarios received by email did not open correctly. Shared scenarios now launch the app, import successfully, and display a confirmation message.
Multiple Screens
Enhanced the code preview screen with improved keyboard visibility, automatic uppercase entry, confirmation before deletion, visible expiry dates, and dynamic list sizing.
Multiple Hardware
Improved firmware update handling during active scenarios. Scenarios now pause automatically when the firmware screen appears and resume afterward. Firmware prompts no longer shut down scenarios or cause black screens, and updates can be completed without losing progress. The “slide to exit” gesture is disabled while the firmware screen is visible to prevent accidental interruption.
Generic Screen
Improved Generic screen HR and PR behavior. HR displays as dashes when ECG is unavailable, automatically switches to PR monitoring, links alarms to PR when HR is off, and flashes PR values when exceeding alarm limits.
iSimulate Community
Resolved an issue where “Import from REALITi” did not respond during media import.
Camera
Corrected an issue causing the Video Module display to appear upside down. Recordings now display with correct orientation, and previewed and saved videos match the live feed. 11.20.0 Mar 1 Atlas
We’ve released a new firmware update for the Atlas Baby device to improve performance and reliability:
Updated firmware to Version 1.5.1 11.19.1 Jan 22 Hamilton T1
Beta features: We have introduced EtCO₂/SpO₂ modelling. Spontaneous Breathing & NIV remain beta and continue to be refined. Interpret the impact of NIV leak on VTE with caution.
High Flow / HiFlowO₂: Improved bar graph auto-scaling for oxygen and flow with independent scaling and smoother transitions. Refined HiFlowO₂ trends with correct right-to-left updates, consistent time scaling, hourly markers, a fixed 6-hour Trends view, and resolved Standby dial and monitoring update issues.
Ventilation monitoring: Corrected monitored IE, Ti, Te, and VTe to display real-time values. Fixed an incorrect VTE alarm message in NIV mode.
Controls & access: Ventilator controls, modelling control, and the ventilator tab are now visible and active for all users. NIV, NIV-ST, and HiFlow modes are enabled by default.
Waveforms & pressure: Corrected pressure display and behaviour in SIMV+, PSIMV+, SPONT, NIV, and NIV-ST modes where pressure support is present, including on additional monitors. Fixed waveform and numeric distortion when multiple T1 monitors are connected.
Waveform scaling & markers: Pressure, Flow, and Volume waveform scales now adjust accurately across their full range at the correct timeline point. Fixed trigger marker alignment and improved spontaneous breath markers to display at the start of each breath cycle.
Dials & parameters: Improved EMV responsiveness once VTE and fTotal stabilise. Refined orange dot indicators to correctly represent soft min/max limits. Improved dial alignment, interaction, and value handling across adult and neonatal modes, including correct carry-over or reset when switching modes or patient types.
Standby & sync: Improved Standby control behaviour for Rate, Vt, and ΔPcontrol. Corrected Plimit/PEEP relationship rules and improved PSync behaviour for synchronised and independent adjustments.
Event handling & stability: Prevented parameters from auto-updating or appearing while editing events in REALITi Control, and stopped PaO₂/FiO₂ Ratio from being automatically added when creating new events while connected to a T1 screen. Fixed crashes when switching from Hamilton T1 to other monitors, including scenarios using 18-lead ECG with ventilators.
Spontaneous breathing logic: Corrected spontaneous breath detection during ventilation by fixing EtCO₂ set-point logic.
SimOccluder
REALITi integrated with SimOccluder
Simconnect 2
Added support for SimConnect 2
Facilitator
Fixed Advanced Respiratory Controls RR alignment on iPad (9th Gen)
Corrected HiFlowO₂ flow display range when connected to T1
Improved ECG behaviour after shock so post-shock signals update immediately
Fixed Pacer Settings “+” button visibility on 10.2″ iPads (German translation)
Updated SVV behaviour (5–50% range, default 5%, all options visible, randomiser when RR = 0)
Added detailed connection logs with timestamps
Fixed Occlusion layout issues on newer iPads (iOS 18+)
Defibrillators & Monitors
Zoll X: Fixed filtered CPR ECG disappearing after shock
Zoll X Advanced: Added support for configurable escalating defibrillation energy
Nihon Kohden EMS1052: Fixed pacing button labels, CPR text/timer visibility, and waveform/message layering issues
Corpuls: Fixed AED Analyse button flashing and corrected Auto Limits alarm behaviour
TMS: Restored scenario switching from the student data area
Scenarios, Screens & Hardware
Fixed scenario sharing between iPads on the same Wi-Fi network
Prevented accidental closures caused by iOS 26.x gestures on multiple screens
Added battery level indicators for CPR Band and CPR Puck (charge status not supported)
Atlas: Improved CO₂ waveform behaviour during ventilation, enabled etCO₂ display from first breath, and fixed a Wi-Fi configuration crash on Atlas Junior
Simplified CPR Band renaming 11.19.0 Jan 6 Atlas
• Improved rhythm display accuracy on Atlas Junior and Baby by ignoring slight compressions (≤10 mm Junior, ≤6 mm Baby).
• Resolved PSIMV mode issues: PSync now saves correctly, pInsp changes send properly, and switching PSync updates all controls.
• Added ventilator settings to Scenario Events: parameters can be viewed, edited, and synced directly from event cards.
Camera
• Enhanced NIBP display and logging: values now appear in exported videos, redundant logs removed, and event labels simplified.
Corpuls
• Improved pacing display: C3 shows correct pacing lozenge; C1 applies continuous therapy with visible ‘STIM’ label.
• Fixed Scenario Review Timeline for accurate playback.
• Updated German translations: “Cancel” → “Abbr.”, “Demand” in uppercase, “Schrittmacher” → “Schrittm.”
CPR Band / CPR Puck
• Added Bluetooth signal strength (RSSI) logging to detect interference.
• Logs now show “OK” or “Warning” based on signal stability.
Hamilton T1
• Aligned Facilitator settings with Hamilton T1 parameters for consistent operation.
• Adjusted ranges, increments, and validations for all key controls.
• Improved spontaneous rate and I:E ratio interaction for stable timing.
• Standby timer now starts only after a ventilation cycle.
• Alarm settings retain correct values when switching modes or patient types.
• Neonatal alarms reset to defaults and display valid limits.
• Resolved overlapping Pressure Alarm dials to maintain proper top/bottom order.
• Improved Trends tab and submenu behavior for smoother navigation.
• Menus now close automatically when entering Standby.
• Fixed I:E ratio display in SIMV+ and PSIMV+ modes.
• Ventilation and ExpMinVol values display correctly with one decimal and retain settings after weight or alarm changes.
• Adjusted PEEP/CPAP dial so max indicator appears only within valid range.
• Patient type switching now resets values correctly; “Last Patient” retains defaults.
• Resolved alarm dial overlaps for proper display order.
Nihon Kohden – EMS1052
• Fixed CPR elapsed time reset and “Restart CPR” timing (now 10 s).
• Improved Japanese Cardioversion messages with corrected layout and font.
• Added suspend time messages: “Suspend 10s,” “20s,” and “Restart CPR 30s” appear during pauses and clear on resume.
Scenarios
• Added 100+ training scenarios: 73 Adult, 17 Paediatric, 13 Neonatal.
• Organized under new folder structure; existing scenarios unchanged.
Multiple Hardware
• Fixed “No connected device” message display after disconnecting all devices.
• Added automatic Bluetooth reconnection after unintended disconnects.
• Manual disconnects remain user-controlled.
• Renamed devices retain connections with consistent log updates.
Multiple Screens
• Fixed behavior when electrodes or pads are disconnected.
• Heartbeat sounds and rate alarms now stay inactive until proper connection. 11.18.0 11/25/2025 Atlas
CPR feedback now displays correctly on supported additional monitors when the primary monitor is connected to a CPR device.
Resolved crashes that occurred when selecting CPR Console as an additional monitor while the primary monitor was connected to a CPR device.
Fixed crashes during compressions on Schiller Defigard and Physiogard monitors when they were used as additional displays.
Corrected CPR logging on Atlas Adult Gen 1 and Gen 2 manikins so recoil depth values now record accurately instead of always showing “-1.”
Corpuls
Corrected DEMAND mode message when ECG electrodes are removed; the message now properly states “No ECG cable in DEMAND mode” instead of “Pause.”
Fixed AED text and voice prompts when switching from Monitor mode so they remain consistent with the pie chart state.
Restored the “Perform CPR” voice prompt when returning to AED with the pie chart running.
Adjusted Start Auto Analysis so AED prompts.
Resolved an issue where AED voice prompts could still be heard while the device was in Monitor mode.
Replaced “Up” and “Down” text buttons with arrow icons on Pacing intensity and frequency screens, and improved how adjustments respond to arrow key input.
Updated arrow icons in soft key buttons, the Settings menu, and the Manual Defib energy selection screens for clearer consistency across the interface.
Improved German translations for Advanced ICP Controls, Advanced PAP Controls, and multiple clinical or monitoring terms.
Added missing translations for Imported Image and Imported Video labels.
Corrected casing for the USE OFFLINE button and resolved translation inconsistencies for Urine – Ketones.
Standardized terminology where existing German translations were unclear or inconsistent.
CPR Band
Added detailed logging for CPR band activity, making it easier to identify issues and improve troubleshooting during use.
Facilitator
Updated default ETCO₂ to 39 and respiratory rate (RR) to 12 in both Control and scenarios.
Updated the REALITi Welcome Scenario and quick start scenarios to use these new values (changes apply if “copy default scenarios on startup” is enabled and the app is restarted; any custom scenarios remain unaffected).
Nihon Kohden – EMS1052
Added translations for the 12-Lead ECG button and front panel in Japanese, Spanish, French, German, Portuguese, Italian, Polish, and Romanian.
Adjusted layout to ensure all translated text displays correctly.
Updated AED and Pacing label styling for better visibility.
Fixed a display issue in Manual mode where the compression graph incorrectly appeared during energy charge when no CPR device was connected.
Corrected HR value display so numerics no longer shrink when highlighted in Large and Normal views, and restricted highlighting when the ECG source is PAD.
Improved CPR assist visuals: bars now show the correct starting compression depth, bar width and height have been refined, and grey bars were added for clearer release and compression feedback.
“CPR Good Compressions” now displays on a white background, with turquoise bars when compressions meet correct depth and rate, while other bars remain red.
Improved spacing for compression values, “CPR-Comp,” and “Tempo /m” labels to ensure consistent readability.
Fixed ETCO₂ display in Control when connected to EMS1052: removed the incorrect third “pale” eye state, leaving only OFF (grey with line) and ON (blue without line).
Toggling the ETCO₂ eye now correctly updates waveform and text visibility for both ETCO₂ and RR together.
Lifepak 15
Joule settings for cardioversion shocks now remain at the last manually entered value (for example, 100 J) instead of auto-escalating for subsequent shocks.
Maintenance
Upgraded the Instabug iOS SDK to version 16.0.2 to maintain stability and compatibility with upcoming iOS releases.
Zoll X Advanced
Heart rate and pulse rate alarm parameters now adjust correctly when switching patient mode to Pediatric or Neonate. 11.17.0 10/13/2025 Atlas
Pulse settings now remain consistent across manikins, even after disconnection or monitor changes.
Age profiles replace manikin type for clearer management: Adult, Paediatric, and Neonate set directly in the scenario panel.
Alarm limits and CPR settings sync automatically with the selected age profile.
Facilitator panels act only as remote controls, with the monitor managing and sharing settings.
Corpuls
Pacing mode in C1 improved: Mode button active, selectable Demand/Fix modes, clear prompts, and full German translations.
Added confirmation prompts for pausing, resuming, and switching off pacing, with German translations.
Electrode handling in C1 improved: ECG waveform now disappears when pads are disconnected, correct prompts are displayed, and switching leads in defibrillator mode is more reliable.
Pacing clarity enhanced: added warning beeps, flashing alarm lights, clear on-screen alerts, refined Demand/Fix behavior, and full German translations for all pacing messages.
Simplified settings by removing unsupported escalation protocol on C1 and C3.
Fixed ECG curve issues: curves remain consistent when changing leads or modes, reset correctly on power-off, and unnecessary options removed.
Fixed audio issues in Pacer mode on C3: removed incorrect prompts and metronome carryover.
Top ECG curve in AED/Manual modes on C3 now fixed to ECG only, with lower curves freely configurable.
Fixed alarm activation in C1 for vital signs outside limits, improving reliability of messages and sounds.
Fixed C1 Send button so CPR and analysis are no longer interrupted.
Facilitator / Control
Improved stability with auto-reconnect, clearer connection lost messages, and better handling of disconnect/reconnect scenarios.
Hamilton T1
PEEP and Plimit handling improved: correct limits, alarms, and display markers.
TI dial increments corrected: +0.01 up to 1.0, then +0.05 after.
Spont Set Point default set to 40, with proper adjustment controls.
Added Respiratory Set Point to simulate spontaneous breathing response to CO₂ levels.
Spontaneous effort now stops if ETCO₂ falls below the set point.
Added quick buttons for adjusting both Spont Effort and Set Point.
Corrected Flow Trigger dial limits for Adult/Pediatric and Neonatal modes.
Fixed Flow Trigger display so the Off option appears correctly depending on the mode.
Waveforms and numerics now update instantly instead of only once per breath.
PEEP changes now display correctly during both inspiration and expiration.
Pressure-controlled modes now respond more accurately to PEEP adjustments.
Flow trigger now checked in real time, with spontaneous breath markers appearing at the true trigger point.
Nihon Kohden - EMS1052
Fixed ECG2 and CPR waveform display issues, ensuring proper alignment on smaller screens.
Updated Japanese translations across CPR messages, vital signs, alarms, and patient info.
Improved CPR feedback: added “CPR Good Compressions” message, refined depth/rate prompts, and alternating guidance.
CPR bars and feedback now restricted in Child mode, with clear messages shown.
Added CPR assist waveforms and graphs, including compression depth/tempo bars and CPR numerics in Monitor mode.
Expanded translations for AED, Manual, and Pacing modes, patient info, shock messages, and pacing prompts.
Improved overall Japanese language coverage and accuracy.
Translation Improvements
Added Romanian language support.
Zoll
Fixed a crash when opening Patient Info on Zoll X screens used as additional monitors. 11.16.0 09/17/2025 Atlas
* Fixed Atlas Adult Wifi Config showing wrong manikin type.
* Fixed wifi settings not updating for Atlas Adult Gen3
* Fixed wifi config not disconnecting after restart manikin.
* Fixed not able to open wifi config settings for Atlas Adult Gen3
* Fixed Atlas Adult Gen3 compression depth mapping table 11.15.1 08/07/2025 Atlas
Fixed manikins on old firmware not getting correct type
Additional checking of manikin types
Depth parameter changes with multiple CPR devices attached
Fixed timeout
Increased timeout to 7 seconds for Atlas Baby, Junior, AdultGen3
Fixed pulse level setting not reflecting pulse level
CPR artefacts appear on ECG trace when using Atlas
SpO2 waveform changes to Poor Perfusion when ventilation or compression is started
Camera
Fixed waveforms not running on camera screen
Corpuls
Removed energy escalation protocol settings (C1, C3)
Improved C1 invert display
Stopwatch responds to scenario pauses and resumes
Fixed C1 QRS tone disappearing when in either defibrillator or pacer modes and C3 QRS tone disappearing when switching from home screen to defib modes
Analyse and Charge buttons activate AED and Manual Defibrillator modes, respectively
Fixed CPR button showing no artefacts on the Pacing rhythm
Pacer off prompt appears while pacing is on when changing from Monitor mode to AED or Manual defibrillator
Resolved “Connect pacer cable” appearing in C1
Removed middle soft key option at the start of AED including during CPR phase and second soft key option ‘Manual’ to ‘Energie’ in Manual mode
Fixed visible soft key options when charging or charged in Manual mode to only have Energy, Cancel and Monitor in both EN and DE
Resolved Metronome setting text in AED and Manual modes
Resolved pacing rhythm in Demand mode when only pads are on
Extra checks on RA or LL electrodes as part of the pacing rhythm state to ensure it affects pacing rhythm
Corrected pacing rhythm not affected by pads when off
Improved how the rhythm with pacing reacts when either electrodes or pads are removed and then re-attached
Fixed text prompt for Pacing when switching to defib
Fixed ‘Switch off Pacer?' transitions when the switch is cancelled while electrodes are off
Added translation ‘'Schrittmacher ausschalten?’
Fixed C1 audio doubling voice prompts
Improved C1 Views
Facilitator/Control
Fixed primary folder appearing empty when it’s not
Added multiple lines for scenario names
Generic
Fixed Sync not getting disabled after a shock
Hamilton T1
Removed ‘Connect Ventilator’ button
Fixed default Delta P value
Fixed spontaneous breath marker not being removed at start and end of trace
Spont TI is automatically calculated
Improved calculations of default TI when changing Adult IBW or Neonatal Weight
Improved IBW/weight calculation for the Vt alarm limit
Improved the SPONT mode by disabling the option and setting value to 0 in Ventilator settings
Fixed Pressure cmH2O alarm limits on high and low for Neonatal and Adult
Improved alarm ExpMinVol default limits’ calculation from weight/IBW, Vt and Rate
Corrected ExpMinVol low and high limit values for each patient type
Fixed ExpMinVol low and high limits where it skips the maximum limit value
Fixed fTotal high and low limits where shows Adult limits instead of Neonatal
Fixed Vt limit low value skipping value 10 before the Off symbol for Adult
Fixed Vt limit high value to show maximum of 3000 for Adult before the Off symbol
FixedVt limit high value to show maximum of 300 for Neonatal before the Off symbol
Fixed Vt limit low value to show maximum of 0.1 for Neonatal before the Off symbol
Improved Vt limit low value calculation based on weight or IBW
Fixed Alarms limits' where the high value can get lower than the low value
Improved Rate and TI values
Fixed dial subtext
Improved TI max dot
Added indicator for Nebulizer when it’s active
Added message “Nebulizer is active”
Created 10-sec timer for the generic message display
Added a stop action for the Nebulizer to remove message and turn off light indicator
Nihon Kohden - EMS1052
Improved Japanese translations
Phillips MRX
Display CPR waveform on all leads when CPR is active
Teaching Management System (TMS)
Fixed data computation, depth guide, and compression target rendering
Zoll X Advanced
Adjusted label sizes so translations fit appropriately 11.15.0 07/31/2025 Camera
Fixed waveforms not running on the camera screen.
Atlas
Fixed CPR artefacts not appearing on ECG trace when using Atlas. 11.14.1 06/22/2025 Atlas
Added support for Atlas Adult Next Gen
Implemented Atlas Baby bootloader firmware update. It will only update if there is a new firmware MCU app update
Corpuls
Corrected texts to “corpuls simulation”
Corrected "VT/VF possible" translation in DE for C3
Improved metronome automatic start settings for manual defibrillator and AED, to display correct metronome setting when switching between AED and manual defibrillator
Set the on state default value to Adult 30:2 for the automatic start of the metronome in both manual and AED settings
CPR Band
Updated CPR recommendation statement for CPR device renaming
Updated renaming rule to accept only the new naming format
Facilitator/Control
Added a randomiser for the Arterial line to be in effect when the line is inserted while RR value is zero.
Resolved crashing when ‘HOLD TO SPEAK’ is tapped while recording
Added a message in the User Recordings window to indicate when no recordings are saved
Disabled the 'HOLD TO SPEAK' button during recording or until the recording is saved or discarded
Added ‘HOLD TO SPEAK’ button in dark mode
Updated UI for ‘New Recording’ menu when recording
Added prevention on selecting other Sounds categories while recording
Changed how User Recordings and New Recording appear, selecting User Recordings now will show both together
Hamilton T1
Resolved the issue of Neonate weight dial minimum value lowering the minimum value of Vt past the limit.
Set up the default value for parameter Rate to depend on IBW (adult) or weight (neonatal) values
Adjusted Neo SPONT mode limits to Basic, More and Apnea options
Improved Neo SPONT mode values
Resolved the issue of the Standby Screen becoming unresponsive after pressing the Power button while a parameter dial is active (orange)
Activated Tools window button
Initialised Tools window
Added initial Utilities and Configuration tabs including contents
Updated Neonatal UI options in Standby mode
Added the three Neonatal quick setup buttons
Created the Neonatal dial with default of 2.0kg and allow weight adjustment
Improved Last patient tab UI with ability to recall the last patient data
Corrected the default Last Patient tab on first load of Standby
Initialised Events window
Nihon Kohden - EMS1052
Created a simulated CPR waveform
Added CPR waveform in AED, Manual and Monitor modes (comp bars excluded)
Added CPR elapsed time including how it resets when a new CPR device connection is established
Resolved Pad messages appearing behind the CPR guide lines
Updated the JA text display ‘Do 5 more chest compressions. Do not touch patient’ in AED by removing an unnecessary symbol
Resolved the issue of multiple texts appearing on Press Scale options while in JA language
Resolved the issue of ‘Use Disposable Pads’ voice overlap with ‘Press AED button’
Added restriction to the use of the charge button when Pads are off
Resolved the issue of “CPR Assist” message appearing in Pacing even when there is no CPR assist connected
Corrected the EN text display ‘Do 5 more chest compressions. Do not touch patient’ in AED by removing an unnecessary symbol
Resolved the issue of waveforms (EtCO2, SpO2) resetting after Send is pressed from Control
Initialised behaviour for Alarm slider’s up and down arrows when it reaches the limit by changing it to a light grey colour when they become ‘disabled’ or ‘off’
Zoll
Resolved where Zoll monitors were experiencing crashes after switching from a different monitor with VT/VF alarms occurring before or after the switch
Added 8-sec timer in Zoll X monitors to close defibrillator window and return to home screen after a shock is delivered
Multiple Screens
Resolved parameter values for vitals: HR, SpO2, ETCO2, etc. not updating correctly
Translation Improvements
Updated Brazilian Portuguese translations
Added adjustments to Zoll X, LifePak 15, and CPR Console to ensure Brazilian Portuguese translations fit appropriately
Updated Simplified Chinese REALITi Translation 11.14.0 06/11/2025 Facilitator/Control
* Added hiding of subfolders in slideout scenario list on control and scenario new start
Menu / main screen
* Renamed 'Stage' to 'Monitor' in the main menu
Multiple Screens
* Fixed sounds muting when opening the app from the background
* Fixed waveforms becoming laggy when doing a UI action causing the QRS sound to be incorrect timing for HR. This mainly affected older iPads
* Added 0.5 second delay for shock button presses on all defibrillator when sync is on
Translation Improvements
* Added Brazilian Portuguese translation corrections
Corpuls
* Updated C1 AED/Manual voice and text prompts for EN and DE
* Stop media players in sound managers when checking therapy electrodes
* Change the float menu for defib to exclude some options
* Changing the rotary knob when it’s in AED
* Added German translations
* Add VF/VT alarm in Corpuls3
* Modify VF/VT in defib mode to not show
Hamilton T1
* Fixed the Rate dial not increasing in PCV+ mode
* Added Print Screen for
* Battery indicator and icon improvements
* Added the Screen Lock feature to Hamilton T1
Nihon Kohden - EMS1052
* Removed unnecessary Large Numerics' text displayed in waveform line
* Resolved lead waveforms lines not showing correctly when disconnected
* Improved CPR Assist CPR messages
* Fix issue Press menu blank
* Fix issue in Large Numeric
* Technical and guidance messages improvements
* Fixed crash related to translations in the Patient information area preventing startup.
* Removed the Temp1 Check Sensor message on start when starts without cable plugged in to match the real machine.
* Added miscellaneous improvements for the parameter alarm sliders and window
Zoll R
* Resolved the issue of Pacer spikes not appearing on the Zoll R screen 11.13.0 05/21/2025 Atlas
Handled manikin ages for the following monitors CPR feedback dashboard: Zoll X Series/Advanced/Propaq MD, Zoll R Series, Generic Defib, Schiller Defigard 7. CPR depth ranges can be customized on control
Handle age changes for center compression feedback and compression depth
Corpuls
Fixed a bug on the C3/C3T where Demand mode was possible with no electrodes attached
Re-added 'Stimulation fortsetzen' to C3 Pacing
Resolved the issue of 'Start Analysis Automatically' setting from C3 affecting C1 text prompts display
Added correct text translation “PreShock CPR durchführen” for C3
Corrected text prompts ‘Analyse starten’ and ‘Start Analysis’ for C1
Improved Pacer pause and continue pacing prompts
Amended Pacer pause and continue pacing DE translations
Resolved the issue of being unable to charge after powering on with the manual button
Resolved the issue of overlapping DE voice prompts while the “Patient nicht berühren” prompt was playing
Resolved the issue of prompting the user to start analysis when it was not required after a shock delivery during the switch from Manual to AED
Resolved the issue where the pie chart gets reset incorrectly during a switch from Manual to AED (vice versa) while charging; the pie chart reset will now only occur if it is fully charged
Facilitator/Control
Resolved the issue of ART/BP line showing zero values in Generic screens
Added support for multiple scenario folder levels
Hamilton T1
Added a ‘Feature not Simulated’ popup to the O2 enrichment key for Hamilton T1
Added the day/night toggle for Hamilton T1
Lifepak 15
Added preShock CPR user input in LIFEPAK 15 Settings
Added preShock CPR in LifePak15
Added preShock CPR duration restrictions: values 0-14 are considered Off state, values 15-30 activate preShock, and values above 30 are capped at 30
Set preShock CPR duration default at 15s
Implemented printing out of waveforms currently displayed on the screen
Resolved the issue of ‘Shock Advised’ audio overlap
Resolved the issue of “Analyzing now--Stand clear” text abruptly flashing when Analyze button is pressed during a CPR countdown
Nihon Kohden - EMS1052
Default lead sources in different modes improvement
Added CPR assist linkage icons
Created connectivity status for connected and disconnected CPR device
Added temporary static CPR-Comp image for CPR device connections for AED mode only
Fixed screen reloading when switching between modes
Resolved the issues of 12-Lead ECG waveforms not appearing properly in other smaller iPad screens
Improved translations for the Nihon Kohden EMS 1052
Resolved the issue on Lead change in 12-Lead ECG where it displayed incorrect and incomplete leads
Improved Lead changes when switching to 12-Lead ECG where it will change source Pad to II and other leads will remain as is
Corrected lead source when switching from Large Numerics to 12-Lead ECG 11.12.0 04/14/2025 Atlas
* Informs the user if wifi change is successful or not for Atlas Baby
* Changed the WiFi change result message to be multiline
* Added support for extended mac so the WiFi scan works with the new firmware in Atlas Baby
* Fixed message in Atlas Baby about new update when on WiFi not showing
* Fixed disconnect issue at end of update in Atlas Baby
* Improved update speed in Atlas Baby
* Added firmware update for Atlas Baby
* Updated Atlas Baby firmware to 1.4.8
Corpuls
* Resolved the issue on cardioversion settings where Autosync is not set as default setting in both C1 and C3
* Corrected order of mode selection to Async, Sync and Auto Sync in both C1 and C3
* Fixed C1 - Double up of voice prompt to press analyse
* Fixed C1 - No oximetry pulse tone
* Corrected translation “Schrittmacher ausschalten?”
* Added the missing text after a shock is performed in Manual defibrillator, ‘Shock performed - Perform CPR’ in both EN and DE
* Changed pie chart symbol to shock (lightning) symbol in C1
* Added pause and resume features on AED/Manual defibrillator’s elapsed and pie chart timers of C1 and C3 for scenario pauses
* Applied pause and resume features on C1 and C3 pie charts for scenario pauses
* Created timer reset features on C1 and C3 power buttons
* Resolved the issue on AED/Manual defibrillator timers not resetting when switching between monitors from Control
Nihon Kohden - EMS1052
* Resolved the issues on overlapping alarm messages of IP disconnections.
* Added Zero CAL message limitations - messages will disappear when calibration is done
* Added STAT mode
* Assigned STAT as default when Patient type is changed
* Initialised oscillation graph images to NIBP process - generic images only
* Added cancellation of NIBP calculation when Manual is ready to shock or when AED analysis has started
* Fixed small glitch on rendering dash line affecting some monitors
* Resolved the issue of incorrect paper positioning on different screen sizes
* Fixed the issue where the paper comes out from a different area when scrolling is used
* Added second tab in Press menu - alarms to be improved in later tickets
* Created a priority of display for CO2 and ART line since REALITi supports more than two sockets - other Critical Care added will not be displayed unless one of the two is disconnected
* Resolved the issue of the ‘More Setup’ button toggles the IP tab when clicked
* Resolved the issue of ART showing up in Press even when disconnected
* Corrected Press current tab highlight(blue) on selection
* Added Zero Cal for secondary IP attachments
* Added Press Label for second socket
* Allowed enabling back of used Press Label when an IP is disconnected
* Improved the More Setup menu for the second label, with correct tab repositioning when disconnected
* Improved IP label removal in the More Setup menu when the priority CO2 connection is established
LIFEPAK 35
* Resolved issues on the bottom menu not displaying correctly after a shock in AED
Multiple Screens
* Fixed fluctuating HR on slower iPads
Zoll X
* Changed EtCO2 to be off by default 11.11.0 03/19/2025 Atlas
* Corrected icon for Atlas Baby Wifi Config.
* Confirm dialog added to ask before changing to the firmware update screen.
* Applied random variation to the ventilated volume.
* Set no reply on pulse command for Atlas Baby/Junior.
Corpuls
* Resolved the issue of missing text and voice prompts in C1.
* Updated translations
* Added C1 text translations for ‘Ready for charging’, ‘Manual Defibrillation Mode?', ‘Press Analyse’ and 'Analysing heart rhythm’
* Added C3 text DE translations for the AED and Manual
* Updated QRS/PT mute icon
* Added ‘PT’ text to show as QRS/PT
* Resolved the issues on configurations not retained after changing scenarios or windows.
* Improved function on softkey “Views” where it will reset configuration back to default including selecting the current view.
* Added voice file for “Assess heart rhythm
* Added symbols on certain texts in both AED and Manual
* Changed voice and text from “Start analysis to “Press analyse” (AED)
* Changed voice “Do not touch or move patient” to “Do not touch patient, analysing heart rhythm” (AED)
* Changed text “Do not touch or move patient” to “Do not touch patient” (AED)
* Changed text from “Shock performed” to “Shock performed, perform CPR” (AED and Manual)
* Lessened playback delay on voice prompt “Perform CPR” to appear like a continuous voice prompt after a shock is performed (AED)
* Changed text “Select Energy” to “Press Analyse” with voice prompt. (AED)
* Changed text “Ready for shock” to “Press and hold down Shock key”(Manual)
* Changed text “Select Energy” to “Select energy / Charge energy” (Manual)
* Changed text “Select Energy” to “Assess heart rhythm” with the voice prompt when switching to Manual or when pie chart is full.
* Corrected text for when shock is cancelled or CPR is continued to “Select energy / Charge energy” (Manual)
* Added DE voice file prompts.
* Added restrictions on Analyze and Shock buttons while HR eye button from Control is off for both Corpuls AED and Corpuls 3.
* Applied the proper prompts when HR eye is off for both Corpuls AED and Corpuls 3
* Resolved the issue on Corpuls AED Analyse key flashing even though the Pads from Control is off.
* Resolved the issue on both Corpuls AED and Corpuls 3 not recognising the state of eye from Control when switching between these screens.
Lifepak 20
* Resolved the issue of crashing when the speed dial selects Patient’s menu from Options.
Nihon Kohden - EMS1052
* Introduced support for the Nihon Kohden - EMS1052
Scenarios
* Enabled scenario version 3 for web build of control.
Ventilator Engine
* Added a more realistic IE ratio to etco2 waveform.
Zoll
* Resolved the issue on Zoll monitors crashing (Zoll R not affected) when a CPR device is connected, and made improvements to other lines of codes.
Zoll R
* Resolved the issue on sync dots appearing in the second waveform instead of the first waveform.
Zoll X
* Resolved the crash issue on Zoll X when SimConnect and a CPR device are connected.
* Fixed the most visible issues on UI overlaps on CPR and BVM windows. 11.10.0 03/03/2025 ATLAS
* Changed good compression rate from 90-120 to 100-120.
* Streamlined the source code for CPR depth.
* Resolved the issue on CPR depth values incorrectly displaying rounded numbers.
* Fixed the issue on CPR Adequate depth line overlapping CPR Maximum depth line.
* Fixed depth label cutoff in Zoll monitors.
* Added pulse and HR to Atlas device settings.
Corpuls
* Updated Crowdin translations to include “Schocktaste gedrückt halten” and “R-EKG Vermessen”.
* Corrected encoding to UTF-8
* Added “Connect therapy electrodes” audio prompts for both EN and DE.
* Corrected DE audio files for “schock abgeben” and “schock abgegeben”
* Changed timer symbol to pie chart symbol.
* Corrected messages and voice audio prompts for “Do not touch patient. Analyzing heart rhythm” when the Analyze button is pressed.
* Added new German voice audio prompts.
* Added “Press Analyze” message and voice audio prompt when pie chart is filled.
* Corrected translation for “Measuring D-ECG”.
* Enabled ECG print on Print button.
* Added restriction on Print button if D-ECG preview is on.
* Added the missing Cancel button when shock is ready in AED.
* Corrected effect of Cancel button on CPR pie chart.
* Resolved the issue on the Perform CPR not getting activated after a shock disarms automatically.
Zoll (incl. Zoll X)
* Resolved the issue on Zoll monitors crashing (Zoll R not affected) when a CPR device is connected, and made improvements to other lines of codes.
* Resolved the crash issue on Zoll X when SimConnect and a CPR device are connected.
* Fixed the most visible issues on UI overlaps on CPR and BVM windows.
Multiple Screens
* Resolved the issue on the NIBP SEND where it toggles ART line when pressed.
* Set Zoll ETCO2 connection as ON always without the need to toggle CO2 button.
* Fixed the issue on Schiller Defigard/Physiogard Touch 7 and Tempus Pro where the SpO2 connection is not recognised when switching to these monitors.
* Resolved the issue on LifePak20 SpO2 pulse bar not appearing on connection.
* Corrected the LifePak20 “SpO2: Searching for Pulse” message to appear only when there is no finger connection.
* Resolved the issue on monitors not showing the correct state of SpO2 when switching from a ventilator.
* Resolved the issue of app crashing when swiping from the left to exit Mindray D3/D6 or Nihon Kohden EMS1052 while the AED is in process.
* Added extra checks to prevent similar crashes in Mindray D3/D6, EMS1052 and LifePak35.
Translation Improvements
* Added translations for Polish, Estonian, Czech, Slovak, Korean.
* Corrected translation for “Measuring D-ECG”. 11.9.0 01/22/2025 Scenarios
- Fixed the issue on textbox not getting added in a new Scenario.
- Resolved the issues on custom scenarios not loading properly. 11.8.1 01/08/2025 Atlas
* Updated Atlas Junior compression depth mapping
* Changed disconnect timeout fix for Atlas Junior to > V2
* Fix for better CPR artifact in Atlas Junior
* Enabled firmware update to latest on connect for production
* Fixed Atlas Junior/Baby not processing data stream after packets arrive too fast or not enough data in the packet
* Added support for old Atlas Junior firmware
* Enabled Atlas Junior firmware update for production build
* Updated Atlas Baby’s adequate and maximum depth values
* Fixed clearing of Atlas Junior/Baby wifi config from available list on scan
* Rounding volume for Atlas Baby, removed random variation of volume for Baby/Junior
* Reduced Junior firmware upload timeout
* Increased wait time after Junior firmware upload before requesting restart
* Added more compression idle checks for Junior/Baby
* Added reset CPR on no data from Junior/Baby manikin
* Added checks in Junior/Baby to reduce issue of abnormally high rate
Corpuls
* Modify defibDisarmed for AED to make it aedCprDidBegin every disarming
* Added CPR Depth changes handler in C1
* Fixed adequate and max depth not reverting to original values after disconnecting manikin
* Fixed overlapping lines and truncating labels
* Fixed displaying “Deliver Shock” when it was unnecessary
* Fixed Auto Analyse where it does not automatically activate after pie chart cycle
* Fixed metronome turning off
* Fixed staying in “Select Energy”
* Improvements on metronome state during switches to and from AED
* Set metronome default for both Manual and AED as Adult 30:2
* Adjusted the metronome after preShock to ensure the sound is correct for “Adult cont” or “Child cont”
* Resolved voice prompt activating randomly during CPR phase after exiting and restarting the same monitor
* Resolved incorrect buttons (Energy, 100, 200) appearing in AED
SimConnect
* Resolved additional monitor not showing the correct state of pleth based on the primary monitor
* Fixed removal of available SimConnects on rescan when bluetooth disabled
* Enabled SimConnect firmware update for production build
* Fixed monitors crashing when attempting to connect to SimConnect
Scenarios
* ECG, Spo2, Etco2 and RR default visible state changed to off for new scenarios and existing scenarios keep existing behaviour of being on
* Changed finger attached to be on by default
Facilitator Control
* Fixed selected Events not loading after update on new electrodes
* Added 18 channel configuration for custom Event
* Enhanced Electrode Placement to update based on monitor, displaying 12 or 18 channels
* Changed Quick Placement and Manual Placement
* Enabled switching between Anterior and Posterior views
* Enabled new electrodes in Manual Placement and Quick Placement
* Adjusted UI to be suitable for smaller screens
Defigard Touch 7
* Resolved Defigard Touch 7 crashing when electrode placement is changed
* Fixed crash from removing 12 leads and pads before scenario starts in Defigard Touch 7
Lifepak 20
* Resolved memory leaks that slow performance
Lifepak 15
* Added ART option in Channel 2 and Channel 3 for waveform
* Fixed Co2 numeric not blinking when outside of alarm range
Zoll X/Advanced
* Set Pads as default trace
* Added Pads as an option in the Quick Access Key: Lead
TMS
* Resolved exported video not getting the correct RR and Temperature values
* Resolved ETCO2 and RR showing inaccurate values during preview
* Fixed NIBP getting hidden when changing parameters
Translations
* Updated The Netherlands translations
CPR Depth Monitor
* Resolved monitor with CPR device crashing upon connection with Control
CPR Ventilation Monitor
* Added scaling on ventilation bars and numeric label for max volume based on manikin type or parameter settings
* Included a line for each max volume and set at 90% of the full scale
* Enabled clipping of values greater than the full scale
Multiple Hardware
* Added disconnecting all connected devices on closing app 11.8.0 12/16/2024 **New Features**
CPR Ventilation Monitor
* Added scaling on ventilation bars based on manikin type or parameter settings.
* Added numeric label for max volume based on manikin type or parameter settings.
* Incl. a line for each max volume and set at 90% of the full scale.
* Enabled clipping of values greater than full scale.
CPR Depth Monitor
* Added CPR depth configuration from Control on Zoll X Series, Zoll X Advanced, Zoll Propaq MD, Corpuls 3, Zoll R Series, Generic Defib and CPR Console.
* Enabled conversion of units for compression depths.
* Improved CPR bar displays for compression depths.
* Adjusted CPR bars to match CPR depth values.
* Fixed monitor with CPR device crashing upon connection with Control.
CPR Depth Control / Facilitator
* Added CPR depth configuration in Parameter Settings
* Added conversion of units for Adequate Depth.
* Added conversion of units for Max. Depth.
* Added CPR manikin type drop down option for Automatic, Newborn, Child and Adult.
CPR Depth
* Fixed compression depth values not updating properly when switching to another monitor.
* Fixed issue on adequate compression depth value not displayed in the correct level.
* Fixed monitor crashing when switching to other monitors while Atlas Junior is connected.
* Corrected Corpuls3 and Corpuls 3T values showing zeros instead of the default values.
**Improvements**
Atlas
* Increased length of low RR CO2 when using SimConnect or Atlas manikin for ventilation.
* Fixed Atlas CPR not affecting HR.
* Added manikin data recording (Currently debug builds only)
* Added CPR depth configuration from Control on Zoll X Series, Zoll X Advanced, Zoll Propaq MD, Corpuls 3, Zoll R Series, Generic Defib and CPR Console. Enabled conversion of units for compression depths. Improved CPR bar displays for compression depths. Adjusted CPR bars to match CPR depth values.
* Added partial support for Atlas Baby.
* Fixed AED not working from “movement detected” due to compressions state not ending.
* Made connect process for Atlas Junior same as in 11.6.0.
* Cap data of Atlas Junior.
* Added extra values in Atlas Junior compression lookup table.
Corpuls 1
* Fixed Hamilton T1 where the alarm continues after quitting a scenario.
* Corrected therapy socket influence on electrodes.
Corpuls 3/3T
* Fixed issue with metronome count not reverting to original configured ratio.
* Fixed issue with Start Analysis being played just after powering on, no delay.
* Fixed issue with metronome not resetting to Adult 30:2 when C3/C3T is powered off.
* Fixed memory leak in Corpuls 3/3T which resulted in a crash when used with Hamilton T1.
Defigard Touch 7
* Added parameter field flash on NIBP, Temp, ETCO2 and RR when values are off limits.
* Enabled flash on bigger NIBP field.
* Improved priority alarms to show in the alarm status field properly.
* Corrected precedence of alarm message from any parameter over “ECG: Cable not detected” message.
* Added “ECG: VF/VT” if HR alarm still persists for a certain period of time for any patient type.
* Included “ECG: Asystole” message when HR is 0.
* Improved alarm light to change from orange to red if alarm is for VF, VT or Asystole.
* Improved alarm light to show orange for medium and low alarms.
* Fixed Pulse field not flashing when pulse is off limits.
* Fixed “SpO2: Pulse off limits” message issue from not appearing when there is a pulse rate alarm.
* Integrated pulse alarm limits to the alarms; value on the pulse will now determine whether to show the alarm based on these limits.
Hamilton T1
* Fixed Hamilton T1 where the alarm continues after quitting a scenario.
Multiple Monitors / Screens
* Fixed the additional monitor not showing the correct state of pleth based on the primary monitor.
* Fixed ECG waveform not being changed if the new waveform was the same as the previously loaded scenario.
Zoll R
* Fixed Filtered ECG not getting deactivated by Control. 11.7.0 12/09/2024 Multiple Screens
- Improved logs window performance when lots of logs.
- Enabled unit measurements' conversion in IBP and NIBP of the following monitors: Zoll R, Philips MRx, Schiller Defigard Touch7, Generic Monitor, Generic Defib, Generic Wave, Generic Numerics, GE Carescape B850, Philips IntelliVue MX800 and WelchAllyn Connex.
- Monitors that do not support the conversion to unit cmH2O will remain as mmHG.
- Resolved issue in Schiller Defigard Touch7 where NIBP value gets changed by the conversion toggle button in CO2 in Unit Measurements
- Enabled Pacer Settings' Default Output and Rate for Zoll Propaq MD, Zoll X, Schiller Defigard Touch 7 and Tempus LS.
- Enabled Pacer Settings' Remember Last Output and Rate for Zoll Propaq MD, Zoll R, Zoll X, Schiller Defigard Touch 7 and Tempus LS.
- Fixed ECG waveform not being changed if the new waveform was the same as the previously loaded scenario.
Capnostream 35
- Resolved crash issue on starting Capnostream 35.
Corpuls 3/3T
- Fixed issue with metronome count not reverting to original configured ratio.
- Fixed issue with Start Analysis being played just after powering on, no delay.
- Fixed issue with metronome not resetting to Adult 30:2 when C3/C3T is powered off.
Hamilton T1
- Changed F Total b/min from decimal to integer in Alarms.
- Changed F Total b/min from decimal to integer in the Monitoring section.
Lifepak 15
- Resolved issues on NIBP countdown timer and MAP display overlap.
- Enabled reset on contrast when the monitor is turned off.
- Resolved issues with the Home and high-contrast Sunvue buttons, where the lower portion was unresponsive to clicks.
- Resolved crash issue when changing back from high contrast Sunvue display to color display.
- Resolved issue on metronome.
- Enabled correct state for CPR button when switching from AED to manual defibrillator.
Revel Vent
- Resolved issue on audible tone looping after swiping from left to exit the ventilator.
Schiller Defigard Touch 7
- Enabled flashing alarm light and parameter field flashes in the secondary Pulse field when limits are - exceeded.
- Enabled alarm light flash for HR.
- Added heart icon in the second Pulse parameter field
- Fixed primary Pulse background to blue when changing from HR to Pulse.
- Resolved issue on secondary Pulse not changing to dashes when Control’s finger probe is off.
- Resolved issue on HR, SpO2 and Pulse not flashing in both alarm light and parameter fields when limits are - exceeded.
- Fixed issue on primary Pulse not changing back to HR field when the source is AUTO and becomes unhidden.
Zoll R
- Resolved issue on “NIBP MEAS ABORTED” message appearing when NIBP value in Control Parameter Settings is changed. 11.6.0 10/21/2024 Logs
- Added ventilation logs to the log window. (was previously in dev only generic ventilator).
Facilitator
- Fixed Ventilator tab heading going over the top of the Monitor popup.
- Added a switch in the Ventilator tab and Advanced Respiratory controls to disable the ventilation modeling for SpO2 and EtCO2. RR will still reflect what is selected on the ventilator.
Additional Monitors
- Fixed crash caused by reconnecting to the monitor.
Schiller Defigard/Touch 7
- Corrected waveform fields display including ECG labels when pads or leads are OFF.
- Corrected the differences between HR Source Auto and specific ECG Curves List sources to display their respective sources.
- Fixed ECG waveform visibility conflicts with SpO2 and ETCO2.
- Fixed bug on Auto Scale that incorrectly changes label of selected source.
- Fixed bug on Amplitude that incorrectly changes label of selected source.
- Corrected HR source display when Pads is OFF.
- Corrected display for first waveform field to show no line when disconnected from Control.
Corpuls 3
- Fixed CPR prompt audio duplication.
- Fixed volume icon's display for both presses and dial turns.
- Added yellow bars for deep compressions.
- Added CO2 as second waveform in AED/Manual mode.
Corpuls 1
- Fixed out of memory crash. 11.5.5 08/29/2024 New Features:
- No new features.
Improvements:
Translations
- Updated German translations.
Facilitator
- Fixed flatlined pleth in preview.
Schiller Defigard/Touch 7
- Corrected Control connections on electrodes and pads with HR source: Auto and Defi.
- Corrected waveform label display for HR source.
- Enabled prioritising ECG source when HR source set in Auto.
- Corrected display values for HR and Pulse parameters.
- Added ECG messages on connection and disconnection.
- Added SpO2 messages on connection and disconnection.
- Fixed HR measurement field to stay visible even when disconnected.
- Corrected HR value when disconnected.
- Enabled heart symbol in HR measurement field.
- Enabled SpO2 module in Pulse parameter.
- Added SpO2 sound level.
- Fixed HR and SpO2 sound source including parameter display.
- Removed SpO2 sound changes pitch in settings.
Lifepak 15
- Added Temperature parameter.
- Enabled connection to Temperature in Control.
- Enabled ability to change units in Control’s Parameter Settings.
- Added Temperature parameter in Speed dial selection.
Zoll X-series/R-series
- Fixed shock button with Sync releases shock.
- Fixed missing charge level view when shock is disarmed.
Zoll Propaq M/Propaq MD/X-series
- Enabled message and progress indicator for NIBP Auto.
Mindray D3
- Fixed yellow alarm 3 tones every 20 seconds. 11.5.4 08/07/2024 Bug fixes and performance enhancements 11.5.3 07/18/2024 Bug fixes and performance enhancements 11.5.2 07/04/2024 Bug fixes and performance enhancements 11.5.1 06/26/2024
REALITi version 11.16.0
Updated At: 2025-09-17
Changes: Sep 17, 2025
Version 11.16.0
Atlas
Pulse settings now remain consistent across manikins, even after disconnection or monitor changes.
Age profiles replace manikin type for clearer management: Adult, Paediatric, and Neonate set directly in the scenario panel.
Alarm limits and CPR settings sync automatically with the selected age profile.
Facilitator panels act only as remote controls, with the monitor managing and sharing settings.
Corpuls
Pacing mode in C1 improved: Mode button active, selectable Demand/Fix modes, clear prompts, and full German translations.
Added confirmation prompts for pausing, resuming, and switching off pacing, with German translations.
Electrode handling in C1 improved: ECG waveform now disappears when pads are disconnected, correct prompts are displayed, and switching leads in defibrillator mode is more reliable.
Pacing clarity enhanced: added warning beeps, flashing alarm lights, clear on-screen alerts, refined Demand/Fix behavior, and full German translations for all pacing messages.
Simplified settings by removing unsupported escalation protocol on C1 and C3.
Fixed ECG curve issues: curves remain consistent when changing leads or modes, reset correctly on power-off, and unnecessary options removed.
Fixed audio issues in Pacer mode on C3: removed incorrect prompts and metronome carryover.
Top ECG curve in AED/Manual modes on C3 now fixed to ECG only, with lower curves freely configurable.
Fixed alarm activation in C1 for vital signs outside limits, improving reliability of messages and sounds.
Fixed C1 Send button so CPR and analysis are no longer interrupted.
Facilitator / Control
Improved stability with auto-reconnect, clearer connection lost messages, and better handling of disconnect/reconnect scenarios.
Hamilton T1
PEEP and Plimit handling improved: correct limits, alarms, and display markers.
TI dial increments corrected: +0.01 up to 1.0, then +0.05 after.
Spont Set Point default set to 40, with proper adjustment controls.
Added Respiratory Set Point to simulate spontaneous breathing response to CO₂ levels.
Spontaneous effort now stops if ETCO₂ falls below the set point.
Added quick buttons for adjusting both Spont Effort and Set Point.
Corrected Flow Trigger dial limits for Adult/Pediatric and Neonatal modes.
Fixed Flow Trigger display so the Off option appears correctly depending on the mode.
Waveforms and numerics now update instantly instead of only once per breath.
PEEP changes now display correctly during both inspiration and expiration.
Pressure-controlled modes now respond more accurately to PEEP adjustments.
Flow trigger now checked in real time, with spontaneous breath markers appearing at the true trigger point.
Nihon Kohden - EMS1052
Fixed ECG2 and CPR waveform display issues, ensuring proper alignment on smaller screens.
Updated Japanese translations across CPR messages, vital signs, alarms, and patient info.
Improved CPR feedback: added “CPR Good Compressions” message, refined depth/rate prompts, and alternating guidance.
CPR bars and feedback now restricted in Child mode, with clear messages shown.
Added CPR assist waveforms and graphs, including compression depth/tempo bars and CPR numerics in Monitor mode.
Expanded translations for AED, Manual, and Pacing modes, patient info, shock messages, and pacing prompts.
Improved overall Japanese language coverage and accuracy.
Translation Improvements
Added Romanian language support.
Zoll
Fixed a crash when opening Patient Info on Zoll X screens used as additional monitors.