@gkasprow I've observed a small bistability on Booster's power detectors a few times.
I'm running Booster from Urukul at a constant power, frequency and phase. I'm logging all the Booster diagnostics using the VCP interface. The observation is that sometimes the power changes by ~0.3dB and the noise changes at the same time. The only correlation I can spot in the diagnostics is that the 5V MP voltage changes a bit at the same time.
See this screenshot:

@gkasprow can you think of a cause of this??
Next I'll add an external power meter to log the output power and see if this is an issue with the Booster diagnostics or if it's actually on the output...
@gkasprow I've observed a small bistability on Booster's power detectors a few times.
I'm running Booster from Urukul at a constant power, frequency and phase. I'm logging all the Booster diagnostics using the VCP interface. The observation is that sometimes the power changes by ~0.3dB and the noise changes at the same time. The only correlation I can spot in the diagnostics is that the 5V MP voltage changes a bit at the same time.