There seems to be discrepancy in EDIS data import in PLDs and labeling duration to what is provided and what is the reality in different subjects - and it differs acorss subjects.

See image of PLD 1500, labeling duration 82 cycles * 18.4 ms = 1508ms. Which can be detected from DICOM/Phoenix/sWiPMemBlock:

See also #1240

image

sWipMemBlock.adFree[1] = 110
2 = 800000 PLD
3 = 82 NumRFCycles
4 = 360 or 560
10 = 10
11 = 100

Release notes

Calculate Siemens labeling duration from NumRFBlocks.