Notifications
You must be signed in to change notification settings - Fork 13
Closed
Description
Description
There seems to be discrepancy in EDIS data import in PLDs and labeling duration to what is provided and what is the reality in different subjects - and it differs acorss subjects.
See image of PLD 1500, labeling duration 82 cycles * 18.4 ms = 1508ms. Which can be detected from DICOM/Phoenix/sWiPMemBlock:
See also #1240
sWipMemBlock.adFree[1] = 110
2 = 800000 PLD
3 = 82 NumRFCycles
4 = 360 or 560
10 = 10
11 = 100
Release notes
Calculate Siemens labeling duration from NumRFBlocks.
Metadata
Metadata
Assignees
Labels
bugSomething isn't workingSomething isn't working